SOM CODE:1890/107
MODEL :DCP 3000
COMPANY: Vacuubrand Germany
1890/107_DCP 3000
Description
DESCRIPTION
The fully electronic vacuum measuring device DCP 3000 is both versatile and decisive with its mechanical robustness, measuring accuracy and reliability. The large, illuminated display provides the measured value display of all connected sensors via an easy-to-use rotary control. Up to eight vacuum sensors (four of each type VSK 3000 / VACUU·VIEW and VSP 3000 / VACUU·VIEW extended) can be connected to the DCP 3000. The measurement can be carried out directly on the application. The DCP 3000 enables a relative pressure measurement (reference sensor VSK 3000) as well as working with pressure-dependent switching points by specifying switch-on and switch-off points on an external valve. The vacuum measuring device can also be used as a data logger with up to 32,000 measuring points. The data is read out via the serial interface (RS 232C). The vacuum sensor VSP 3000 is based on the gas thermal conductivity (Pirani) principle for the wide measuring range from atmospheric pressure to 10-3 mbar. Thanks to its structure made of chemically resistant plastics (PBT, PUR) and ceramics, it offers significantly higher chemical resistance and robustness compared to conventional Pirani sensors with fine wire metal filaments. Please also note our alternative product VACUU·VIEW extended.
Technical data
Upper measuring limit in mbar/hPa | 1000 mbar | |
Upper measuring limit in torr | 750 torr | |
Lower measuring limit in mbar/hPa | 1 x 10-3 | |
Lower measuring limit in torr | 1 x 10-3 torr | |
Measurement principle | Thermal conductivity (Pirani), chemically resistant plastics / ceramics | |
Accuracy of measurement | +- 15% of indicated value in the range 0.01-5 mbar/hPa/torr mbar/hPa/torr | |
Vacum connection | Small flange KF DN 16 and hose nozzle DN 6/10 mm | |
Lower ambient temperature (storage) | -10 °C | |
Upper ambient temperature (storage) | 60 °C | |
Lower ambient temperature (operation) | 10 °C | |
Upper ambient temperature (operation) | 40 °C | |
Max. media temperature for continuous operation | 40 °C | |
Max. media temperature short term | 80 °C | |
Material of outer housing | Robust plastic housing with good chemical resistance | |
Protection class | IP 20 | |
Protection class frontside | IP 42 | |
Weight | 0.44 kg | |
Dimension L in mm | 144 mm | |
Dimension W in mm | 124 mm | |
Dimension H in mm | 114 mm | |
Vacuum sensor | VSP 3000 | |
Cable length of vacuum sensor | 2 m | |
Cable length connection in m | 2 m | |
NRTL certification | Yes | |
Items supplied | Vacuum gauge with external vacuum sensor VSP 3000, ready for operation, with wall-plug power supply and manual. |
Reviews
There are no reviews yet.