4004/40_BSC-1500IIB2-X

4004/40_BSC-1500IIB2-X

SafetyCabinet/LaminarFlow/PurifierBench

BIOBASE BIOTECH Company mainly provides Class II biological safety cabinets that provide personnel, environmental and product protection. Both room air and interior cabinet air are drawn into a front grille creating an air barrier that provides personnel protection

SOM CODE:4004/40
MODEL :BSC-1500IIB2-X
COMPANY: JINAN BIOBAS,CHINA

 

Description

DESCRIPTION

I: Operating Range
1.1 Running conditions 110V~220V / 50Hz ~ 60Hz
1.2 Operating temperature range: 15-35
1.3 Operating humidity range ≤85%
II Parameter and Function
2.1 Filter
2.1.1 Inflow filter: HEPA filter,
2.1.2 Filter efficiency: 99.999%(0.1-0.2μm)
2.1.3 Level of exhaust filter: HEPA filter
BSC-1500IIB2-X
Introduction
Welcome to BIOBASE
2.1.4 Exhaust filter efficiency: 99.99%(0.3μm)
2.1.5 Display of negative pressure and filter life: Bar Code
2.1.6 Display of accumulation working hours: VFD
2.1.7 Filter replacement: front of the cabinet
2.2 Electric Motor
2.2.1 Frequency conversion electric machine could make aural and visual alarm by the
blockage state of the filter
2.2.2 Automatic adjustment of the rotating speed to keep the wind speed.
2.3 Airflow System
2.3.1 Air Circulation: 0%
2.3.2 Exhaust Air: 100%
2.3.3 Inflow Velocity: 0.65 ± 0.15m/s
2.3.4 Down flow Velocity: 0.37±0.015m/s
2.4 Front Window
2.4.1 Sensitive Alarm: The height of the front window could adjust at will, when overtop
the safety height, it will alarm
2.4.2 Safety Glass: toughened glass, explosive-Proof and prevent UV
2.4.3 Thickness of front panel is more than 5mm
2.5 Workroom Interior
2.5.1 Material: Table-board is made of excellent stainless steel,two flanks and back side
are punched into a smooth whole body, more easy to cleaning.
2.5.2 Socket: 2 internal waterproof socket
2.5.3 Lighting >800lux
Welcome to BIOBASE
2.5.4 UV Lamp: Sterilizing by UV lamp, the UV light couldn’t turn on when the
fluorescent lamp turning on or the front window open
2.5.5 Air Grill: Air will be drawn into the cabinet through the air grill.
2.5.6 Water tight socket: Two single water tight sockets are provided on the sidewall for
small devices or equipment used within the cabinet.
2.5.6 Power Socket.
2.6 Size
2.6.1 internal size(w *d *h): 1500x850x2200mm
2.6.2 external size(w *d *h) : 1300x600x660mm
2.7 Noise: <65 dB
2.8 Power: 0.45Kw
2.9 Weight: 180Kg
2.10 Vibration: The net displacement is no more than 3μm when the frequency is from
10Hz to 10kHz
2.11 Stability: The core part of work surface can stand more than 23kg pressure
without permanent distortion.
2.12 Biosafety
2.12.1 Personnel safety: Hammerblow samples ≤10CFU/time
Slit samples ≤5CUF/time
2.12.2 Product safety: The microorganism germ falls ≤5CUF/time
2.12.3 Cross infection safety: The microorganism germ falls ≤2CUF/time
III Configure
3.1 Main Unit
Welcome to BIOBASE
3.2 Inflow ULPA
3.3 Exhaust HEPA
3.4 UV lamp
3.5 Light
3.6 Socket
Accessories
4.1 Water Valve
4.2 Gas Valve
4.3 Vacuum Valve
4.4 Removable Stand
4.5 Remote control
Certificate
5.1 Class III medical equipment registration certificate issued by SFDA
5.1 ISO9001 ISO14001 ISO13485 C

Technical Specification

Contact Us

+966-556111951 info@somatco.com

Visit Our Showroom

Find our Showroom/Branch Offices which is near to you

Reviews

There are no reviews yet.

Be the first to review “4004/40_BSC-1500IIB2-X”

Your email address will not be published. Required fields are marked *